ULTIMA MOS 5 HART
12
12-13 Unsigned-16 Event Cause – See device specific table
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined
10.12 Command #148: Read Maintenance Event Log
This command reads the Maintenance Event Log as specified by the event log number. Event 0 is the
most recent event. Event 1 is the one just before that and so forth.
Request Data Bytes
Byte Format Description
None N/A N/A
Response Data Bytes
Byte Format Description
0 Unsigned-8 Event Log Number
0 – 3 Unsigned-32 Event Running Time (in Seconds)
4– 6 Date Event Date: Day, Month, Year – 1900
7 Unsigned-8 Event Hour
8 Unsigned-8 Event Minute
9 Unsigned-8 Event Second
10-11 Unsigned-8 Reserved = 0
12-13 Unsigned-8 Code: 0-Gas check
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined