MSA Ultima MOS-5 Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 31
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
ULTIMA MOS 5 HART
19
10.23 Command #170: Set Current Range
Request Data Bytes
Byte Format Description
0 Unsigned-8 0 – Range 3.5 mA – 20 mA, 1 -- Range 1.25 – 20 mA
Response Data Bytes
Byte Format Description
0 Unsigned-8 0 – Range 3.5 mA – 20 mA, 1 -- Range 1.25 – 20 mA
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1 - 2 Undefined
3 Error Passed Parameter Too Large
4 Undefined
5 Error Too Few Data Bytes Received
6 - 127 Undefined
10.24 Command #189: Set Sensor Life
Request Data Bytes
Byte Format Description
0 Unsigned 8 Sensor life, %
Response Data Bytes
Byte Format Description
0 Unsigned 8 Sensor life, %
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1 – 4 Undefined
5 Error Too Few Data Bytes Received
6 – 127 Undefined
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Comentarios a estos manuales

Sin comentarios