10.19 Command #156: Read Calibration Event Log
This command reads the Calibration Event Log as specified by the event log number. Event 0 is the most
recent event. Event 1 is the one just before that and so forth.
Request Data Bytes
Byte Format Description
None N/A N/A
Response Data Bytes
Byte Format Description
0-3 Unsigned-32 Event Running Time (in Seconds)
4-6 Date Event Date: Day, Month, Year – 1900
7 Unsigned-8 Event Hour
8 Unsigned-8 Event Minute
9 Unsigned-8 Event Second
10 Unsigned-8 1 - N/A, 2 – Calibration
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined
10.20 Command #163: Get Fast Changing Information
This command reads the Fault Event Log as specified by the event log number. Event 0 is the most recent
event. Event 1 is the one just before that and so forth.
Request Data Bytes
Byte Format Description
None N/A N/A
Response Data Bytes
Byte Format Description
0 – 1 Unsigned-16 Mode – depends on instrument
2 – 3 Unsigned-16 Sub Mode – depends on instrument
4 – 7 Float Analog Output
8 – 9 Unsigned-16 Priority Fault
10 – 11 Bit map Error Status