MSA Ultima MOS-5 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 31
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
ULTIMA MOS 5 HART
14
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined
10.15 Command #151: Set Run Time Meter
This command sets the internal run time meter.
Request Data Bytes
Byte Format Description
0 – 3 Unsigned-32 Run Time Meter Value
Response Data Bytes
Byte Format Description
0 – 3 Unsigned-32 Run Time Meter Value
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1 – 4 Undefined
5 Error Too Few Data Bytes Received
6 – 127 Undefined
10.16 Command #152: Read Run Time Meter
This command reads the internal run time meter.
Request Data Bytes
Byte Format Description
0 N/A N/A
Response Data Bytes
Byte Format Description
0 – 3 Unsigned-32 Run Time Meter Value
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 30 31

Comentarios a estos manuales

Sin comentarios