ULTIMA MOS 5 HART
5
8.0 Universal Commands
Command 3 returns the current loop variable and the primary variable for a total of 9 bytes returned.
Command 9 returns the PV only.
9.0 Common Practice Commands
The following common practice commands are implemented.
9.1 Supported Commands
The following common-practice commands shown in Table 3 are implemented:
Command
Number
Byte
Number
Meaning
Command 38 N/A
Reset Configuration Changed Flag
Command 48 0 Returns Priority Fault, High Byte
Command 48 1 Returns Priority Fault, Low Byte
Command 48 2 Returns error status (same as Modbus register x02), High Byte
Command 48 3 Returns error status (same as Modbus register x02), Low Byte
Command 48 4 Returns Power Cycled Flag
Command 48 5 Returns Event Happened Flag
Command 48 6
Returns 0x01 = “Maintenance Required”or 0x02 = Alarm or
Warning
Command 48 7 Returns 0
Table 3: ULTIMA MOS 5 – Common Practice Commands
9.2 Burst Mode
The ULTIMA MOS 5 does not support Burst Mode.
9.3 Catch Device Variable
This ULTIMA MOS 5 does not support Catch Device Variable.